![]() |
Following the divergence correction,
I perform a normal moveout correction (NMO) using the estimated RMS
velocities. A resulting CMP gather can be seen in the upper
panel of Figure
.
The near offsets are at the left side and increase to the right side.
A significant change in shape and
amplitude of the source wavelet is visible at the transition zone between the
100 m and 50 m group spacings. The loss in frequency and amplitude
is caused by the strong hydrophone array attenuation.
In the direction of the far offsets, where the hydrophone groups are shorter,
the wavelet sharpens and regains its large amplitudes. These
waveform shapes and amplitude changes can seriously degrade any subsequent AVO
analysis. In order to regularize the wavelet as much as possible, I
perform a
single trace source wavelet deconvolution before NMO correction which
estimates a different filter for each trace.
Deep water (more than 3 km) and shallow dips in the area cause the
wavefront propagation angles to change only moderately as a function of depth.
For example, in Figure
at offset 2.65 km ( transition between
100 m and 50 m groups spacing), the seafloor reflection propagates at 22
while the BSR reflection propagates at 20
. This small change in
propagation angles makes trace-by-trace deconvolution a valid tool in this
case.
Although the application of single trace, as opposed to surface-consistent,
deconvolution can result in undesired changes in amplitude and waveform, the
Blake Outer Ridge seismic data are of high enough quality that single trace
deconvolution with careful quality control gives very consistent results.
After deconvolution, the data are bandpass filtered to a range
of frequencies where signal is found at all offsets. This
removes spurious deconvolutional high-frequency noise. Then the data are
again NMO corrected.
A resulting CMP gather after deconvolution, bandpass filtering and normal
moveout correction can be seen in the lower panel of Figure
.
The waveform
has been well calibrated along all offsets.
The remaining amplitude loss, especially that induced by the
hydrophone attenuation, will be addressed in the amplitude calibration
section.